Raw to Dicom conversion

14 visualizaciones (últimos 30 días)
Hamza Naeem
Hamza Naeem el 23 de Oct. de 2017
Comentada: Walter Roberson el 26 de Oct. de 2017
Hello
I have written a code to convert .raw medical images to .dcm format. But I am having some error while trying to add some metadata information.
Following is my code:
fid3 = fopen( 'F:\patient\same patientCT.raw','r');
pCT = fread(fid3, [512*512*160,1], 'int16');
pCT_p = reshape(pCT, [512 512 160]);
fclose(fid3);
new_pCT1=permute(pCT_p,[2 1 3]);
for j =1:size(new_pCT1,3)
File=int16(new_pCT1(:,:,j)+1000);
m = -54.7500 -(j-1)*1.25;
metadata.ImagePositionPatient=[-325 ; -325; m];
metadata.SliceLocation=m;
metadata.InStackPositionNumber=j;
metadata.InstanceNumber=j;
dicomwrite(File,['CT_' int2str(j) '.dcm'], metadata, 'CreateMode', 'copy');
end
Please help me to resolve this issue. Thankyou
  7 comentarios
Hamza Naeem
Hamza Naeem el 26 de Oct. de 2017
Editada: Walter Roberson el 26 de Oct. de 2017
Still not working, and now I am having this kind of error:
Warning: The 'SOPClassUID' and 'ObjectType' parameters are ignored when 'CreateMode' is 'copy'.
> In dicomwrite>checkOptionConsistency (line 947)
In dicomwrite (line 193)
In Untitled (line 19)
Error using dicom_copy_IOD>getIOD (line 138)
Missing required attribute (0008,0016) "SOPClassUID"
Error in dicom_copy_IOD (line 26)
IOD_UID = getIOD(metadata, options, dictionary);
Error in dicomwrite>write_message (line 277)
[attrs, status] = dicom_copy_IOD(X, map, ...
Error in dicomwrite (line 208)
[status, options] = write_message(X, filename, map, metadata, options);
Error in Untitled (line 19)
dicomwrite(File,['pCT' int2str(j) '.dcm'], metadata, 'CreateMode' , 'copy', 'ObjectType', 'CT Image Storage')
Walter Roberson
Walter Roberson el 26 de Oct. de 2017
For example for plain CT image storage you might perhaps use SOPClassUID of 1.2.840.10008.5.1.4.1.1.2
... In short, if you are going to create your own metadata then you have a lot to create. It is usually far far easier to take another existing DICOM image of the exact same type that you want to create, and copy its metadata (setting different patient information and dates and a unique instance identifier).